Laser Components Measure Damage Thresholds!

Our damage threshold measuring station is designed according to ISO 21254: For this purpose, we have a laser available with wavelengths of 1064 nm and 532 nm, pulses of 7 ns to 10 ns, and a repetition rate of 10 Hz. On the sample, the damage profile is recorded objectively.
 
Preparing Measurements
Prior to each measurement, a so-called typical pulse is characterized that portrays the temporal and spatial averaging of a pulse sequence. This must meet defined stability criteria to ensure a reproducible measurement process.

Performing Measurements
The first step is to limit the energy density used for measurements, then to shoot 10 positions on the sample at the same energy density with 200 pulses, and finally to identify potential damage using a scattered light detector. The ratio of damaged spots to the number of positions in total results in the probability of damage for this energy density. This procedure increases for additional energy densities in order to obtain a characteristic damage curve. This results in the determination of the coating’s damage threshold.

If you have any questions please contact us
Samuel Thienel
Technical Sales Engineer
Samuel Thienel
LASER COMPONENTS (UK) Ltd.
CM2 7PR Chelmsford Essex

Related News

This news might interest you

LASER COMPONENTS UK - Your competent partner for optical and optoelectronic components in the United Kingdom.

Welcome to LASER COMPONENTS UK Ltd., your expert for photonics components. Each product in our wide range of detectors, laser diodes, laser modules, optics, fiber optics, and more is worth every Pound (£/GBP). Our customized solutions cover all conceivable areas of application: from sensor technology to medical technology. You can reach us here:

Goldlay House, 114 Parkway
Chelmsford Essex CM2 7PR
United Kingdom

Phone: +44 1245 491 499
Email: info(at)lasercomponents.co.uk

Team of experts
You have questions or need our support?
Please call us.
Samuel Thienel
Technical Sales Engineer
Samuel Thienel
LASER COMPONENTS (UK) Ltd.
CM2 7PR Chelmsford Essex
Contact Form
You would like to send us something? You can reach us by phone and by e-mail.

Laser Components

Goldlay House, 114 Parkway
Chelmsford Essex CM2 7PR
United Kingdom

Phone: +44 1245 491 499

You will be redirected
to the Fiber Technology Website ...